RealTime CT
Description
The Abbott RealTime CT assay is an in vitro polymerase chainreaction (PCR) assay for the direct, qualitative detection of theplasmid DNA for C. trachomatis in female endocervical or vaginal swab specimens, male urethral swab specimens, or in male and female urine specimens.
Abbott RealTime CT/NG Performance
| Technology | Multiplex, homogenous real-time PCR | |
| Sensitivity | 320 plasmid copies per 400 ìl sample | |
| Specificity | No cross-reactivity with 111 organisms | |
| Interference | No interference with 26 substances | |
| Internal Control | Added to lysis buffer during extraction and detected with all samples | |
| Collection Device | Abbott multi-Collect Specimen Collection Kit |
Abbott RealTime CT Comparison Study to Comparator Assay
A total of 623 specimens including female endocervical and vaginal swabs,male urethral swabs, and male and female urines were obtained frompatients attending STD clinics. The swab specimens were alwayscollected first with the comparator assay collection kit and then withthe Abbott multi-Collect Specimen Collection Kit.
| Positive Agreement | Negative Agreement |
| Analyte | Specimens (n) | Ratioa | (%) | 95% Cl | Ratiob | (%) | 95% Cl |
| CT | 621c | 116/127 | 91.3 | (85.03, 95.60) | 488/494 | 98.9 | (97.38, 99.55) |
Design
Development Philosophy
Today's clinical molecular diagnostics laboratory must have confidence in the quality of their C.trachomatis patient results. As a result of our real-time PCRdevelopment philosophy, Abbott RealTime CT is engineered to be bothsensitive and specific for the detection of C. trachomatis from asingle sample.
Accurate Detection of CT is Dependent Upon a Combination of:
- Primer and Probe Design
- Sample Purification
- Incorporation of an Internal Control
Order Information
| Product Name | Package Size | Vysis® Order Number | Abbott Order Number |
| RealTime CT Abbott RealTime CT/NG Amplification Reagent Kit | 192 tests (4 packs of 48 tests per pack) | 01L31-91 | 01L31-91 |
| RealTime CT Abbott RealTime CT/NG Control Kit | 8 CT/NG Negative Control, 16 CT/NG Cutoff Control | 02G28-80 | 02G28-80 |
| RealTime CT Abbott RealTime CT/NG Application CD-ROM | 1 each | 01L70 | 01L70 |