RealTime CT/NG

1105_20348
RealTime CT/NG

Description

The Abbott RealTime CT/NG assay is an in vitro polymerase chain reaction (PCR) assay for the direct, qualitative detection of the plasmid DNA for C. trachomatis and the genomic DNA of N. gonorrhoeae in female endocervical or vaginal swab specimens, male urethral swab specimens, or in male and female urine specimens.

Abbott RealTime CT/NG Performance

Technology   Multiplex, homogenous real-time PCR
Sensitivity   320 plasmid copies per 400 ìl sample
Specificity   No cross-reactivity with 111 organisms
Interference   No interference with 26 substances
Internal Control   Added to lysis buffer during extraction and detected with all samples
Collection Device   Abbott multi-Collect Specimen Collection Kit

Abbott RealTime CT/NG Comparison Study to Comparator Assay
A total of 623 specimens including female endocervical and vaginal swabs, male urethral swabs, and male and female urines were obtained from patients attending STD clinics. The swab specimens were always collected first with the comparator assay collection kit and then with the Abbott multi-Collect Specimen Collection Kit.

  Positive Agreement   Negative Agreement
Analyte Specimens (n) Ratioa (%) 95% Cl Ratiob (%) 95% Cl
CT 621c 116/127 91.3 (85.03, 95.60) 488/494 98.9 (97.38, 99.55)
NG 617d 91/93 97.8 (92.45, 99.74) 518/524 98.9

(97.52, 99.58) 

Design

Development Philosophy
Today's clinical molecular diagnostics laboratory must have confidence in the quality of their C.trachomatis and N. gonorrhoeae patient results. As a result of our real-time PCR development philosophy, Abbott RealTime CT/NG is engineered to be both sensitive and specific for the detection of C.trachomatis and N. gonorrhoeae from a single sample.

Accurate Detection of CT or NG is Dependent Upon a Combination of:

  • Primer and Probe Design
  • Sample Purification
  • Incorporation of an Internal Control
Single Stranded Linear Probe Design
CT/NG Target Illustration

Order Information

Product Name Package Size Vysis® Order Number Abbott Order Number
RealTime CT/NG Abbott RealTime CT/NG Amplification Reagent Kit 192 tests (4 packs of 48 tests per pack) 02G28-91 02G28-91
RealTime CT/NG Abbott RealTime CT/NG Control Kit 8 CT/NG Negative Control, 16 CT/NG Cutoff Control 02G28-80 02G28-80
RealTime CT/NG Abbott RealTime CT/NG Application CD-ROM 1 each 01L70 01L70
RealTime CT/NG Abbott multi-Collect Specimen Collection Kit 500 kits (10 boxes of 50 kits per box) 09K12-01 09K12-01
In Vitro Diagnostic Available in the United States.
 
CE Marked